3-22
Pulse Current Measurements
Pulse current digitization
The following discussion explains how to digitize a current waveform. A programming
example at the end of this section demonstrates proper command sequence for pulse current
digitization.
Overall steps for digitization:
1.
Sync up to desired edge for measurement.
2.
After detecting edge, wait for the internal and also any user trigger delay.
3.
Take specified number of readings. The supply synchronizes to only the first reading.
After taking the first reading, the supply no longer synchronizes to the selected edge nor
does it wait for a trigger delay (internal or user trigger delay).
In the pulse current digitization mode, readings are generated approximately every 274
μ
s for
battery channel (490
μ
s for charger channel) and placed in the instrument measurement output
buffer. The 274
μ
s or 490
μ
s time interval is the sum of the integration period, 33
μ
sec, and the
period required to convert this information into a measurement, approximately 241µsec for
battery channel (457
μ
s for charger channel). The instrument initiates the storage and conversion
process for the desired number of iterations, as specified with the AVERAGE command, when
the :TLEVel threshold is exceeded. The message "DIGITIZE" is displayed instead of readings.
The "NO PULSE" message will be displayed if the pulse is not detected. Pulse current
digitization is selected by disabling trigger synchronization:
SENS[1]:PCUR:SYNC <b>
Battery Channel (#1)
SENS2:PCUR:SYNC <b>
Charger Channel (#2)
<b> = OFF
Select pulse current digitization (trigger synchronization disabled).
= ON
Select pulse current measurements (trigger synchronization enabled).
The commands to set the trigger level and trigger delay for pulse current measurements also
apply for pulse current digitization. However the trigger delay can be set up to five seconds.
SENS[1]:PCUR:SYNC:DEL <NRf> Battery Channel (#1)
SENS2:PCUR:SYNC:DEL <NRf> Charger Channel (#2)
<NRf> = 0 to 5
User trigger digitization delay in seconds (10µsec
steps). For digitization, the internal trigger delay is
15
μ
sec.
To detect the pulse, the digitization process synchronizes to the edge specified by the
following command:
SENS[1]:PCUR:MODE <name>
Battery Channel (#1)
SENS2:PCUR:MODE <name>
Charger Channel (#2)
<name> =
HIGH or AVER
Sync up to rising edge of pulse for 1st reading of
digitization.
=
LOW
Sync up to falling edge of pulse for 1st reading of
digitization.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...