4-12
Long Integration Measurements
LONG INTEGRAT #1/#2. (Select LONG INTEGRAT #1 for the battery channel and
LONG INTEGRAT #2 for the charger channel.)
4.
Select and adjust the TRIGGER LEVEL and press ENTER. The unit starts looking for
the rising edge of the pulse (this is regardless of the trigger edge setting). If the trigger
level is still too low or too high, the “LONG INT TRIG NOT DETECTED” message will
be displayed briefly. Note that it may take as long as the timeout value for the message
to appear (see “LONG INT TRIG NOT DETECTED message” for more information).
5.
If the message appeared, repeat step 4 until a valid trigger level is found.
6.
Use the MENU key to back out of the menu structure and display long integration
current measurements.
NOTE
For the battery channel, make sure the trigger level range setting agrees with the
trigger level setting set in Step 4.
LONG INT TRIG NOT DETECTED message
The TRIG NOT DETECTED message is possibly displayed when specific TLEV settings
coupled with specific TLEV ranges have been set and a trigger has not been detected. Refer to
for message preconditions.
See step 2 of the
“Long integration measurement procedure” on page 4-10
for information on
setting the trigger level range. For the charger channel #2, the trigger level range setting is not
user selectable.
Table 4-1
TRIG NOT DETECTED message
TLEV setting
TLEV range
TRIG NOT DETECTED
Message
displayed?
100mA
1A
No (not checked because TLEV
setting does not match TLEV
range)
1A
1A
May appear
100mA
5A
Yes (TLEV setting too low)
1.1A
5A
No (valid setting)
3A
5A
Yes (TLEV too high)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...