14-4
Calibration
Resistor connections
When performing calibration steps that use either the 4
Ω
or 3k
Ω
resistor, connect the
Model 2306 OUTPUT #1 or OUTPUT #2 SENSE leads and DMM test leads as close to the
resistor body as possible.
Resistor considerations
The test resistors should be characterized using the 4-wire ohms function of the DMM
recommended in
to measure the resistance values. Use the measured resistance
values to calculate the actual currents during the calibration procedures.
NOTE
The temperature coefficient and temperature change of the 4
Ω
resistor when passing
current at full load must be low enough so that the change in resistance does not cause
incorrect readings. This is summarized in the following equation:
V
OUT
R
R
Δ
+
(
)
----------------------
25
% of Model 2306 amps specification
<
where: V
OUT
is the Model 2306 output voltage
R is the characterized value of the resistor
Δ
R is the change in resistance caused by heating
Front panel calibration
Use the following procedure to calibrate the Model 2306 from the front panel.
summarizes calibration steps.
NOTE
Calibration must be performed in the following sequence, or an error will occur. To
abort calibration and revert to previous calibration constants at any time during the
procedure, press the MENU key. Follow message prompts to abort calibration. Use
▲
and
▼
keys to select desired response.
Step 1: Prepare the Model 2306 for calibration
1.
Turn on the Model 2306 and the digital multimeter, and allow them to warm up for at
least one hour before performing calibration.
2.
Press the MENU key, then choose CALIBRATE UNIT, and press ENTER. The
instrument will display the date last calibrated:
CALIBRATE UNIT
LAST ON 02/01/2003
3.
Press the
▲
key. The instrument will display the number of times it was calibrated:
CALIBRATE UNIT
TIMES = 1
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...