14-10
Calibration
6.
Using the edit keys, adjust the display to agree with the new DMM voltage reading, then
press ENTER. The unit will prompt for Channel #2 DVM full-scale calibration:
CAL DVM CHAN 2
ALL READY TO DO?
7.
Press ENTER to complete Channel #2 DVM full-scale calibration.
8.
Connect the digital multimeter volts input and characterized 4
Ω
resistor to the
Model 2306 OUTPUT #2 SOURCE terminals using the same general connections
shown in
. Be sure to observe proper polarity ( to DMM INPUT
HI; SOURCE - to INPUT LO).
9.
Be sure the digital multimeter DC volts function and auto-ranging are still selected.
10.
At this point, the unit will prompt for 5A Channel #2 full-scale calibration output:
CAL 5 A CHAN 2
SET 1.90000 A
11.
Using the edit keys, adjust the set value to 1.9000A, then press ENTER. The unit will
prompt you for the DMM reading, which calibrates the Channel #2 5A current limit:
CAL 5 A CHAN 2
READ1 1.90000 A
12.
Note the DMM voltage reading, then calculate the current from that reading and the
actual 4
Ω
resistance value: I = V/R. Adjust the Model 2306 current display value to
agree with the calculated current value, and press ENTER.
13.
The Model 2306 will prompt for another DMM reading, which is used for Channel #2
5A measurement calibration:
CAL 5 A CHAN 2
READ2 1.90000 A
14.
Again, calculate the current from the new DMM reading and 4
Ω
resistor value. Adjust
the 2306 current display reading to agree with the new current, then press ENTER.
15.
Disconnect the 4
Ω
resistor, then connect the 3k
Ω
resistor in its place. (Use the
connections in
except make connections to the OUTPUT #2 terminals.)
16.
Make sure the DMM DC volts function and auto-ranging are still selected.
17.
At this point, the unit will prompt to output approximately 5mA for Channel #2 5mA
range full-scale calibration:
CAL 5 mA CHAN 2
ALL READY TO DO?
18.
Press ENTER to output approximately 5mA. The unit will then prompt you for the
DMM reading:
CAL 5 mA CHAN 2
READ1 4.50000 mA
19.
Note the DMM voltage reading, then calculate the current from that voltage reading and
actual 3k
Ω
resistance value. Adjust the Model 2306 current display value to agree with
that value and press ENTER.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...