Pulse Current Measurements 3-15
Command notes (pulse current measurements)
SENSe[1]:FUNCtion ‘PCURrent’
Applies to battery channel (#1)
SENSe2:FUNCtion ‘PCURrent’
Applies to charger channel (#2)
This parameter name can also be enclosed in single quotes (as shown above).
SENSe[1]
:PCURrent
:FAST
Enable or disable pulse current fast readings.
OFF
:SEARch
Enable or disable pulse current search.
ON
:DETect
Enable or disable pulse current detection mode.
OFF
:TimeOUT
Specify length of timeout: 5ms - 32s incrementing in 1ms.
1 (sec)
SENSe2
SENSe subsystem for Channel #2 (charger channel):
:FUNCtion “PCURrent”
Select pulse current measurement function.
VOLT
:PCURrent
Pulse current configuration:
:AVERage <NRf>
Specify average count:
1 to 100 (pulse current measurements), or
1 to 5000 (pulse current digitization).
1
:MODE <name>
Select measurement mode; HIGH, LOW or AVERage.
HIGH
:TIME
Set integration times:
:AUTO
Integration times set automatically.
:HIGH <NRf>
Specify integration time (in sec) for high pulse measurements; 33.33e-6
to 0.8333.
3.333e-5
:LOW <NRf>
Specify integration time (in sec) for low pulse measurements; 33.33e-6 to
0.8333.
3.333e-5
:AVERage <NRf>
Specify integration time (in sec) for average pulse measurements; 33.33e-
6 to 0.8333.
3.333e-5
:DIGitize <NRf>
Specify integration time (in sec) for digitizing or burst measurements (B10
firmware and later); 33.33e-6 to 0.8333.
3.333e-5
:SYNChronize
Pulse detection triggering:
[:STATe]
Send ON to select pulse current measurements or OFF to select pulse cur-
rent digitization.
ON
:TLEVel <NRf>
Set trigger level in amps: 0.0–5.0
0.0
:DELay <NRf>
Specify trigger delay in seconds:
0.0–0.1 (pulse current measurements) or
0.0–5.0 (pulse current digitization).
0.0
:FAST
Enable or disable pulse current fast readings.
OFF
:SEARch
Enable or disable pulse current search.
ON
:DETect
Enable or disable pulse current detection mode.
OFF
:TimeOUT
Specify length of timeout: 5ms - 32s incrementing in 1ms.
1 (sec)
READ[1]?
Trigger and return one reading for Channel #1 (battery channel).
READ[1]:ARRay?
Trigger an array of readings and return them for Channel #1 (battery
channel).
READ2?
Trigger and return one reading for Channel #2 (charger channel).
READ2:ARRay?
Trigger an array of readings and return them for Channel #2 (charger
channel).
Table 3-2
SCPI commands — pulse current measurements (cont.)
Command
Description
Default
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...