9-4
Common Commands
Send *OPC or *OPC?, separated by a semicolon, on the same line with a query (see Ref. A
). If sent on separate lines, an error occurs (B). *OPC or *OPC? can also be sent on
the same line or a separate line with a command that is not a query (C and D).
Table 9-2
*OPC and *OPC? commands
Ref
*OPC
*OPC?
Comment
A
VOLT?; *OPC
SENS:NPLC?; *OPC?
Valid battery channel command line.
SOUR2:VOLTage?; *OPC
SENS2:NPLC?; *OPC?
Valid charger channel command line.
B
VOLTage?; *OPC
SENS:NPLC?; *OPC?
Not valid—query interrupted error.
SOUR2:VOLT?; *OPC
SENS2:NPLC?; *OPC?
Not valid—query interrupted error.
C
CURR 1; *OPC
SENS:NPLC 5; *OPC?
Valid battery channel command line.
SOUR2:CURR 1; *OPC
SENS2:NPLC 5; *OPC?
Valid charger channel command line.
D
CURR 1; *OPC
SENS:NPLC 5; *OPC?
Valid battery channel command line.
SOUR2:CURR 1; *OPC
SENS2:NPLC 5; *OPC?
Valid charger channel command line.
The first line for each reference applies to the battery channel (channel #1) while the second line applies to the charger channel
(channel #2).
*SAV <NRf> — save
Save present setup in memory
*RCL <NRf> — recall
Return to setup stored in memory
Parameters
0 = Memory
location
0
1 = Memory location 1
2 = Memory location 2
3 = Memory location 3
4 = Memory location 4
NOTE
For the Models 2306-PJ and 2306-VS, the memory location settings available are
from *SAV 0 -*SAV 2 and *RCL 0 -*RCL 2 (*SAV 3 and *SAV 4, and *RCL 3 and
*RCL 4 are not available).
Use the *SAV command to save the present instrument setup configuration in memory for
later recall. Any control affected by *RST can be saved by the *SAV command. The *RCL
command is used to restore the instrument to the saved setup configuration. Five setup
configurations can be saved and recalled. *SAV and *RCL are global commands (not channel
specific). Consequently, when a setup is saved or recalled, both channels are affected.
NOTE
The output is always off when a memory location is recalled.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...