9-2
Common Commands
Overview
Common commands are device commands that are common to all devices on the bus. These
commands are designated and defined by the IEEE-488.2 standard. Common commands are
listed in
. Note that detailed information on the Common Commands to program and
read status registers is provided in Section 8.
Table 9-1
IEEE-488.2 common commands and queries
Mnemonic
Name
Description
*CLS
Clear status
Clears all event registers and error queue.
*ESE <NRf>
Event enable command
Program the standard event enable register.
*ESE?
Event enable query
Read the standard event enable register.
*ESR?
Event status register query Read the standard event enable register and clear it.
*IDN?
Identification query
Returns the manufacturer, model number, serial number, and
firmware revision levels of the unit.
*OPC
Operation complete com-
mand
Set the operation complete bit in the standard event register after
all pending commands have been executed.
*OPC?
Operation complete query
Places an ASCII “1” into the output queue when all pending
selected device operations have been completed.
*RCL <NRf>
Recall command
Returns the power supply to the user-saved setup.
*RST
Reset command
Returns the power supply to the *RST default conditions.
*SAV <NRf>
Save command
Saves the present setup as the user-saved setup.
*SRE <NRf>
Service request enable
command
Programs the service request enable register.
*SRE?
Service request enable
query
Reads the service request enable register.
*STB?
Status byte query
Reads the status byte register.
*TRG[1]
Trigger command
Sends a battery channel (#1) bus trigger to the power supply.
*TRG2
Trigger command
Sends a charger channel (#2) bus trigger to the power supply.
*TST?
Self-test query
Performs a checksum test on ROM and returns the result.
*WAI
Wait-to-continue
command
Wait until all previous commands are executed.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...