![Keithley 2000 Скачать руководство пользователя страница 61](http://html.mh-extra.com/html/keithley/2000/2000_calibration-manual_661840061.webp)
Manufacturing calibration commands
Three calibration steps are only performed at the factory or when the unit has been repaired:
:CALibration:PROTected:AC:STEP14
1V AC at 3Hz
:CALibration:PROTected:AC:STEP15
1V AC at 1kHz
:CALibration:PROTected:DC:STEP0
Rear terminal short circuit
:AC:STEP<14|15>
(CALibration:PROTected:AC:STEP<14|15>)
Purpose
To program individual AC manufacturing calibration steps.
Format
:cal:prot:ac:step14 <Cal_voltage>
:cal:prot:ac:step15 <Cal_frequency>
Parameters
<Cal_voltage> = 1 [1V nominal]
<Cal_frequency> = 1E3 [1kHz nominal]
Description
The :AC:STEP command also programs the two manufacturing AC calibra-
tion steps; <n> represents the calibration step number. The appropriate signal
must be connected to the instrument when programming each step, as sum-
marized in the parameters listed above.
Examples
:CAL:PROT:AC:STEP14 1
Program AC step 14.
:CAL:PROT:AC:STEP15 1E3
Program AC Step 15.
:DC:STEP0
(:CALibration:PROTected:DC:STEP0)
Purpose
To perform rear terminal short-circuit calibration.
Format
:cal:prot:dc:step0
Parameter
None
Description
:STEP0 performs the rear short-circuit calibration step in the manufacturing
calibration procedure. Connect a low-thermal short (Model 8610) to the rear
panel input jacks, and select the rear inputs before sending this command.
Example
:CAL:PROT:DC:STEP0
Perform rear short-circuit calibration.
Calibration Command Reference
3-17
Содержание 2000
Страница 1: ...Model 2000Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 10: ...1 Performance Verification ...
Страница 23: ...2 Calibration ...
Страница 45: ...3 Calibration CommandReference ...
Страница 65: ...A Specifications ...
Страница 76: ...A 12 Specifications ...
Страница 77: ...B Error Messages ...
Страница 80: ...C Calibration Program ...