:STEP4
(:CALibration:PROTected:DC:STEP4)
Purpose
To program the -10V DC comprehensive calibration step.
Format
:cal:prot:dc:step4 <Cal_voltage>
Parameter
<Cal_voltage> = -9 to -11 [V]
Description
:STEP4 programs the -10V DC comprehensive calibration step. The allow-
able range of the calibration voltage parameter is from -9 to -11, but -10 is
recommended for best results.
Example
:CAL:PROT:DC:STEP4 -10
Program -10V step.
:STEP5
(:CALibration:PROTected:DC:STEP5)
Purpose
To program the 100V DC comprehensive calibration step.
Format
:cal:prot:dc:step5 <Cal_voltage>
Parameter
<Cal_voltage> = 90 to 110 [V]
Description
:STEP5 programs the 100V DC comprehensive calibration step. The allow-
able range of the calibration voltage parameter is from 90 to 110, but 100 is
recommended for best results.
Example
:CAL:PROT:DC:STEP5 100
Program 100V step.
:STEP6
(:CALibration:PROTected:DC:STEP6)
Purpose
To program the 1k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step6 <Cal_resistance>
Parameter
<Cal_resistance> = 900 to 1.1E3 [
Ω
]
Description
:STEP6 programs the 1k
Ω
4-wire resistance comprehensive calibration step.
The allowable range of the calibration resistance parameter is from 900 to
1.1E3 but 1E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP6 1E3
Program 1k
Ω
step.
3-12
Calibration Command Reference
Содержание 2000
Страница 1: ...Model 2000Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 10: ...1 Performance Verification ...
Страница 23: ...2 Calibration ...
Страница 45: ...3 Calibration CommandReference ...
Страница 65: ...A Specifications ...
Страница 76: ...A 12 Specifications ...
Страница 77: ...B Error Messages ...
Страница 80: ...C Calibration Program ...