DC voltage calibration
After the front panel short and open procedure, the unit will prompt you for the first DC volt-
age: +10V. Do the following:
1.
Connect the calibrator to the Model 2000 as shown in Figure 2-2. Wait three minutes to
allow for thermal equilibrium before proceeding.
NOTE
Although 4-wire connections are shown, the sense leads are connected and discon-
nected at various points in this procedure by turning calibrator external sense on or
off as appropriate. If your calibrator does not have provisions for turning external
sense on and off, disconnect the sense leads when external sensing is to be turned off,
and connect the sense leads when external sensing is to be turned on.
2.
Set the calibrator to output DC volts, and turn external sense off.
3.
Perform the steps listed in Table 2-3 to complete DC volts calibration. For each calibra-
tion step:
• Set the calibrator to the indicated value, and make sure it is in operate.
• Press the ENTER key to calibrate that step.
• Wait until the Model 2000 finishes each step. (The unit will display the CALIBRAT-
ING message while calibrating.)
2000 MULTIMETER
!
R
Model 2000
5700A Calibrator
Output
HI
Input
LO
Output
LO
Note : Use shielded low-thermal cables to
minimize noise. Enable or disable
calibrator external sense as indicated
in procedure.
Sense HI
Sense LO
Sense HI
Sense LO
Input
HI
Calibration
2-9
Figure 2-2
Calibrator con-
nections for DC
volts and ohms
portion of com-
prehensive cali-
bration
Содержание 2000
Страница 1: ...Model 2000Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 10: ...1 Performance Verification ...
Страница 23: ...2 Calibration ...
Страница 45: ...3 Calibration CommandReference ...
Страница 65: ...A Specifications ...
Страница 76: ...A 12 Specifications ...
Страница 77: ...B Error Messages ...
Страница 80: ...C Calibration Program ...