Table 1-3
ACV reading limits
ACV
Range
Applied AC
voltage
Reading limits (1 year, 18°C-28°C)
1kHz
50kHz
100mV
1V
10V
100V
750V
100.0000mV
1.000000V
10.00000V
100.0000V
700.000V*
99.9100 to 100.0900mV
0.999100 to 1.000900V
9.99100 to 10.00900V
99.9100 to 100.0900V
699.355 to 700.645V
99.8300 to 100.1700mV
0.998300 to 1.001700V
9.98300 to 10.01700V
99.8300 to 100.1700V
698.785 to 701.215V
*If the 5725A amplifier is not available, change the 700V @ 50kHz step to 219V @ 50kHz.
(Reading limits for 219V @ 50kHz = 218.362 to 219.638V.)
Performance Verification
1-9