:STEP10
(CALibration:PROTected:DC:STEP10)
Purpose
To program the 10mA comprehensive calibration step.
Format
:cal:prot:dc:step10 <Cal_current>
Parameter
<Cal_current> = 9E-3 to 11E-3 [A]
Description
:STEP10 programs the 10mA comprehensive calibration step. The allowable
range of the calibration current parameter is from 9E-3 to 11E-3. Use the
10E-3 value whenever possible for best results.
Example
:CAL:PROT:DC:STEP10 10E-3
Program 10mA step.
:STEP11
(CALibration:PROTected:DC:STEP11)
Purpose
To program the 100mA comprehensive calibration step.
Format
:cal:prot:dc:step11 <Cal_current>
Parameter
<Cal_current> = 90E-3 to 110E-3 [A]
Description
:STEP11 programs the 100mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 90E-3 to 110E-3. Use
the 100E-3 value whenever possible for best results.
Example
:CAL:PROT:DC:STEP11 0.1
Program 100mA step.
:STEP12
(CALibration:PROTected:DC:STEP12)
Purpose
To program the 1A comprehensive calibration step.
Format
:cal:prot:dc:step12 <Cal_current>
Parameter
<Cal_current> = 0.9 to 1.1 [A]
Description
:STEP12 programs the 1A comprehensive calibration step. The allowable
range of the calibration current parameter is from 0.9 to 1.1. Use a value of
1 whenever possible for best results.
Example
:CAL:PROT:DC:STEP12 1
Program 1A step.
3-14
Calibration Command Reference
Содержание 2000
Страница 1: ...Model 2000Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 10: ...1 Performance Verification ...
Страница 23: ...2 Calibration ...
Страница 45: ...3 Calibration CommandReference ...
Страница 65: ...A Specifications ...
Страница 76: ...A 12 Specifications ...
Страница 77: ...B Error Messages ...
Страница 80: ...C Calibration Program ...