AC voltage calibration
Follow these steps for AC voltage calibration:
1.
Connect the calibrator to the Model 2000 INPUT HI and LO terminals as shown in Fig-
ure 2-4.
2.
Perform the calibration steps summarized in Table 2-6. For each step:
• Set the calibrator to the indicated value, and make sure the calibrator is in operate.
• Press ENTER to complete each step.
• Wait until the Model 2000 completes each step.
Table 2-6
AC voltage calibration summary
Calibration step
Calibrator voltage, frequency
10mV AC at 1kHz
100mV AC at 1kHz
100mV AC at 50kHz
1V AC at 1kHz
1V AC at 50kHz
10V AC at 1kHz
10V AC at 50kHz
100V AC at 1kHz
100V AC at 50kHz
700V AC at 1kHz
10.00000mV, 1kHz
100.0000mV, 1kHz
100.0000mV, 50kHz
1.000000V, 1kHz
1.000000V, 50kHz
10.00000V, 1kHz
10.00000V, 50kHz
100.0000V, 1kHz
100.0000V, 50kHz
700.000V, 1kHz
2000 MULTIMETER
!
R
Output HI
Input LO
Model 2000
5700A Calibrator
Input HI
Output LO
2-12
Calibration
Figure 2-4
Connections for
AC volts calibra-
tion