Front panel calibration
1.
Connect the low-thermal short to the rear panel input jacks, and select the rear inputs
with the INPUTS switch. Allow three minutes for thermal equilibrium.
2.
Press in and hold the OPEN key while turning on the power.
3.
Press SHIFT then CAL, select RUN, then enter the appropriate calibration code (default:
002000).
4.
Select ALL at the CAL:RUN prompt.
5.
Press ENTER.
6.
Perform the entire front panel comprehensive calibration procedure discussed earlier in
this section.
7.
Connect the synthesizer to the Model 2000 INPUT jacks as shown in Figure 2-5. Select
the front input jacks with the INPUTS switch.
8.
After the last AC current calibration step, the instrument will prompt you to enter 3Hz at
1V RMS and 1kHz. For each prompt:
• Low frequency cal: Set the synthesizer to output a 1V RMS, 3Hz sine wave. Use the
left and right arrow keys, and the range keys to adjust the displayed voltage value to
the value you measured and recorded earlier. Press Enter.
• Frequency cal: Set the synthesizer to output a 1V RMS sine wave at 1kHz. Enter
1.000000kHz, then press ENTER.
9.
Set the calibration dates, then save calibration constants to complete the process.
2000 MULTIMETER
!
R
Model 2000
BNC-to-Dual
Banana Plug
Adapter
50
Ω
BNC Coaxial Cable
Model 3930A or 3940 Synthesizer
Main
Function
Output
Note: Synthesizer output voltage
must be accurately measured.
(See text).
Calibration
2-21
Figure 2-5
Synthesizer con-
nections for man-
ufacturing
calibration
Содержание 2000
Страница 1: ...Model 2000Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 10: ...1 Performance Verification ...
Страница 23: ...2 Calibration ...
Страница 45: ...3 Calibration CommandReference ...
Страница 65: ...A Specifications ...
Страница 76: ...A 12 Specifications ...
Страница 77: ...B Error Messages ...
Страница 80: ...C Calibration Program ...