Measuring resistance
The Model 2000 can make 2-wire and 4-wire resistance measurements from 100µ
Ω
to
120M
Ω
.
Connections
Assuming factory default conditions, the basic procedure is as follows:
1.
Connect test leads to the Model 2000 as follows:
A.
For
Ω
2-wire, connect the test leads to INPUT HI and LO.
B.
For
Ω
4-wire, connect the test leads to INPUT HI and LO, and SENSE
Ω
4 WIRE
HI and LO. Recommended Kelvin test probes include the Keithley Models 5805
and 5806. Either the front or rear inputs can be used; place the INPUTS button in
the appropriate position.
2.
Select the measurement function by pressing
Ω
2 or
Ω
4.
3.
Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE
▲
and
▼
keys to select a
measurement range consistent with the expected resistance.
4.
Connect test leads to the resistance as shown in Figure 2-6.
CAUTION
Do not apply more than 1000V peak between INPUT HI and LO or instru-
ment damage may occur.
5.
Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until a normal reading is displayed. Use the lowest possible range for the best resolution.
6.
Take a reading from the display.
2-24
Basic Measurements
Содержание 2000
Страница 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 13: ...1 General Information...
Страница 19: ...2 Basic Measurements...
Страница 54: ...2 36 Basic Measurements...
Страница 55: ...3 Measurement Options...
Страница 87: ...4 Remote Operation...
Страница 139: ...5 SCPI CommandReference...
Страница 215: ...A Specifications...
Страница 226: ...A 12 Specifications...
Страница 227: ...B Statusand ErrorMessages...
Страница 233: ...C Example Programs...
Страница 246: ...C 14 Example Programs...
Страница 247: ...D Models196 199and 8840A 8842ACommands...
Страница 255: ...E IEEE 488 BusOverview...
Страница 271: ...F IEEE 488andSCPI ConformanceInformation...