5-62
SCPI Command Reference
:POSetup <name> command
:POSetup <name>
:SYSTem:POSetup <name>
Program power-on defaults
Parameters
<name> = RST
Select *RST defaults on power up
PRESet
Select :SYSTem:PRESet defaults on power up
SAV0
Select saved defaults on power up
Query
:POSetup?
Query power-on setup
Description
This command is used to select the power-on defaults. With RST selected,
the instrument powers up to the *RST default conditions. With PRES select-
ed, the instrument powers up to the :SYStem:PRESet default conditions. De-
fault conditions are listed in the SCPI tables (Tables 5-2 through 5-11).
With the SAV0 parameter selected, the instrument powers-on to the setup
that is saved in the specified location using the *SAV command.
:FRSWitch? command
:FRSWitch?
:SYSTem:FRSWitch?
Read INPUTS switch
Description
This query command is used to read the position of the FRONT/REAR IN-
PUTS switch. Switch position code is defined as follows:
1 = Front panel inputs selected
0 = Rear panel inputs selected
Содержание 2000
Страница 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 13: ...1 General Information...
Страница 19: ...2 Basic Measurements...
Страница 54: ...2 36 Basic Measurements...
Страница 55: ...3 Measurement Options...
Страница 87: ...4 Remote Operation...
Страница 139: ...5 SCPI CommandReference...
Страница 215: ...A Specifications...
Страница 226: ...A 12 Specifications...
Страница 227: ...B Statusand ErrorMessages...
Страница 233: ...C Example Programs...
Страница 246: ...C 14 Example Programs...
Страница 247: ...D Models196 199and 8840A 8842ACommands...
Страница 255: ...E IEEE 488 BusOverview...
Страница 271: ...F IEEE 488andSCPI ConformanceInformation...