Keithley Instruments, Inc.
Cleveland, Ohio 44139
DRN.
CKD.
APP.
DATE
DATE
DATE
PART NUMBER
LTR
REVISIONS
APP. DATE
FORM 28777A-SBG
BRUNING 40-21 62198-SBG
SPECIFICATIONS
JMM 8/8/01
Rev. E
SPEC-2000
2000 6½-Digit Multimeter Specifications
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
ACCURACY
1
: ±(% of r % of range), 23°C ±5 °C
Voltage Range
Resolution
Calibration Cycle
3 Hz-10 Hz
10 Hz-20 kHz
20 kHz-50 kHz
50 kHz-100 kHz 100 kHz-300 kHz
100.0000 mV
0.1 µV
1.000000
V
1.0 µV
90 Days
0.35 + 0.03
0.05 + 0.03
0.11 + 0.05
0.60 + 0.08
4 + 0.5
10.00000
V
10 µV
100.0000
V
100 µV
1 Year
0.35 + 0.03
0.06 + 0.03
0.12 + 0.05
0.60 + 0.08
4 + 0.5
750.000
V
1 mV
0.035 + 0.003
0.005 + 0.003
0.006 + 0.005
0.01 + 0.006
0.03 + 0.01
Temperature
Coefficient/°C
8
Current Range
Resolution
Calibration Cycle
3 Hz-10 Hz
10 Hz-5 kHz
1.000000
A
1 µA
90 Day/1 Year
0.30 + 0.04
0.10 + 0.04
3.00000
9
A
10 µA
90 Day/1 Year
0.35 + 0.06
0.15 + 0.06
Temperature
Coefficient/°C
8
0.035 + 0.006
0.015 + 0.006
High Crest Factor Additional Error ±(% of reading)
7
CREST FACTOR:
1–2
2–3
3–4
4–5
ADDITIONAL ERROR:
0.05
0.15
0.30
0.40
AC Operating Characteristics
2
Function
Digits
Readings/s
Rate
Bandwidth
6½
3
2s/reading
SLOW
3 Hz-300kHz
6½
3
1.4
MED
30 Hz-300kHz
6½
4
4.8
MED
30 Hz-300kHz
6½
3
2.2
FAST
300 Hz-300kHz
ACV
(all ranges),
and ACI
(all ranges)
6½
4
35
FAST
300 Hz-300kHz
Additional Low Frequency Errors ±(% of reading)
SLOW
MED
FAST
20Hz - 30Hz
0
0.3
—
30Hz - 50Hz
0
0
—
50Hz - 100Hz
0
0
1.0
100Hz - 200Hz
0
0
0.18
200Hz - 300Hz
0
0
0.10
> 300Hz
0
0
0
AC System Speeds
2, 5
FUNCTION/RANGE CHANGE
6
:
4/s.
AUTORANGE TIME:
<3 s.
ASCII READINGS TO RS-232 (19.2k BAUD)
4
:
50/s.
MAX. INTERNAL TRIGGER RATE
4
:
300/s.
MAX. EXTERNAL TRIGGER RATE
4
:
300/s.
AC General
INPUT IMPEDANCE:
1M
Ω
±2% paralleled by <100pF.
ACV INPUT PROTECTION:
1000Vp.
MAXIMUM DCV:
400V on any ACV range.
ACI INPUT PROTECTION:
3A, 250V fuse.
BURDEN VOLTAGE:
1A Range: <0.3V rms. 3A Range: <1V rms.
SHUNT RESISTOR:
0.1
Ω
on all ACI ranges.
AC CMRR:
>70dB with 1k
Ω
in LO lead.
MAXIMUM CREST FACTOR:
5 at full scale.
VOLT HERTZ PRODUCT:
<8 × 10
7
V·Hz.
OVERRANGE:
120% of range except on 750V and 3A ranges.
AC Notes
1
Specifications are for SLOW rate and sinewave inputs >5% of range.
2
Speeds are for 60 Hz operation using factory default operating
conditions (*RST). Auto zero off, Auto range off, Display off,
includes measurement and binary data transfer out the GPIB.
3
0.01% of step settling error. Trigger delay = 400ms.
4
Trigger delay = 0.
5
DETector:BANDwidth 300, NPLC = 0.01.
6
Maximum useful limit with trigger delay = 175ms.
7
Applies to non-sinewaves >5Hz.
8
Applies to 0°–18°C and 28°–50°C.
9
For signal levels > 2.2A, add additional 0.4% to “of reading”
uncertainty.
Содержание 2000
Страница 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 13: ...1 General Information...
Страница 19: ...2 Basic Measurements...
Страница 54: ...2 36 Basic Measurements...
Страница 55: ...3 Measurement Options...
Страница 87: ...4 Remote Operation...
Страница 139: ...5 SCPI CommandReference...
Страница 215: ...A Specifications...
Страница 226: ...A 12 Specifications...
Страница 227: ...B Statusand ErrorMessages...
Страница 233: ...C Example Programs...
Страница 246: ...C 14 Example Programs...
Страница 247: ...D Models196 199and 8840A 8842ACommands...
Страница 255: ...E IEEE 488 BusOverview...
Страница 271: ...F IEEE 488andSCPI ConformanceInformation...