MEASure command
:MEASure[:<function>]?
<function> = CURRent:AC
AC current
CURRent[:DC]
DC current
VOLTage:AC
AC voltage
VOLTage[:DC]
DC voltage
RESistance
2-wire resistance
FRESistance
4-wire resistance
PERiod
Period
FREQuency
Frequency
TEMPerature
Temperature
DIODe
Diode testing
CONTinuity
Continuity test
Description
This command combines all of the other signal oriented measurement com-
mands to perform a “one-shot” measurement and acquire the reading.
When this command is sent, the following commands execute in the order
that they are presented.
:ABORt:CONFigure:<function>:READ?
When :ABORt is executed, if continuous initiation is disabled, the instru-
ment goes into the idle state. If continuous initiation is enabled, the operation
re-starts at the beginning of the Trigger Model.
When :CONFigure is executed, the instrument goes into a “one-shot” mea-
surement mode. See :CONFigure for more details.
When :READ? is executed, its operations will then be performed. In general,
another :ABORt is performed, then an :INITiate, and finally a FETCh? to ac-
quire the reading. See :READ? for more details.
5-6
SCPI Command Reference
Содержание 2000
Страница 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 13: ...1 General Information...
Страница 19: ...2 Basic Measurements...
Страница 54: ...2 36 Basic Measurements...
Страница 55: ...3 Measurement Options...
Страница 87: ...4 Remote Operation...
Страница 139: ...5 SCPI CommandReference...
Страница 215: ...A Specifications...
Страница 226: ...A 12 Specifications...
Страница 227: ...B Statusand ErrorMessages...
Страница 233: ...C Example Programs...
Страница 246: ...C 14 Example Programs...
Страница 247: ...D Models196 199and 8840A 8842ACommands...
Страница 255: ...E IEEE 488 BusOverview...
Страница 271: ...F IEEE 488andSCPI ConformanceInformation...