:AUTO <b>
[:SENSe[1]]:CURRent:AC:RANGe:AUTO <b>
Control auto range for ACI
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO <b>
Control auto range for DCI
[:SENSe[1]]:VOLTage:AC:RANGe:AUTO <b>
Control auto range for ACV
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO <b>
Control auto range for DCV
[:SENSe[1]]:RESistance:RANGe:AUTO <b>
Control auto range for
Ω
2
[:SENSe[1]]:FRESistance:RANGe:AUTO <b>
Control auto range for
Ω
4
Parameters
<b> =
1 or ON
Enable auto range
0 or OFF
Disable auto range
Query
:AUTO?
Query auto range (on or off)
Description
These commands are used to control auto ranging. With auto ranging en-
abled, the instrument automatically goes to the most sensitive range to per-
form the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that
manually selects the measurement range (:RANGe <n>). When auto range
is enabled, the parameter value for :RANGe <n> changes to the automati-
cally selected range value. Thus, when auto range is disabled, the instrument
remains at the automatically selected range. When a valid :RANGe <n>
command is sent, auto ranging disables.
5-42
SCPI Command Reference
Содержание 2000
Страница 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 13: ...1 General Information...
Страница 19: ...2 Basic Measurements...
Страница 54: ...2 36 Basic Measurements...
Страница 55: ...3 Measurement Options...
Страница 87: ...4 Remote Operation...
Страница 139: ...5 SCPI CommandReference...
Страница 215: ...A Specifications...
Страница 226: ...A 12 Specifications...
Страница 227: ...B Statusand ErrorMessages...
Страница 233: ...C Example Programs...
Страница 246: ...C 14 Example Programs...
Страница 247: ...D Models196 199and 8840A 8842ACommands...
Страница 255: ...E IEEE 488 BusOverview...
Страница 271: ...F IEEE 488andSCPI ConformanceInformation...