Introduction
This section contains general information about the Model 2000 Multimeter. The information
is organized as follows:
•
Feature overview
•
Warranty information
•
Manual addenda
•
Safety symbols and terms
•
Specifications
•
Inspection
•
Options and accessories
If you have any questions after reviewing this information, please contact your local
Keithley representative or call one of our Applications Engineers at 1-800-348-3735 (U.S.
and Canada only). Worldwide phone numbers are listed at the front of this manual.
Feature overview
The Model 2000 is a 6
½
-digit high-performance digital multimeter. It has 0.002% 90-day
basic DC voltage accuracy and 0.008% 90-day basic resistance accuracy. At 6
½
digits, the mul-
timeter delivers 50 triggered readings/sec over the IEEE-488 bus. At 4
½
digits, it can read up to
2000 readings/sec into its internal buffer. The Model 2000 has broad measurement ranges:
•
DC voltage from 0.1
µ
V to 1000V.
•
AC (RMS) voltage from 0.1
µ
V to 750V, 1000V peak.
•
DC current from 10nA to 3A.
•
AC (RMS) current from 1
µ
A to 3A.
•
Two and four-wire resistance from 100µ
Ω
to 120M
Ω
.
•
Frequency from 3Hz to 500kHz.
•
Thermocouple temperature from -200°C to +1372°C.
Some additional capabilities of the Model 2000 include:
•
Full range of functions — In addition to those listed above, the Model 2000 functions
include period, dB, dBm, continuity, diode testing, mX+b, and percent.
•
Optional scanning — For internal scanning, options include the Model 2000-SCAN, a
10-channel, general-purpose card, and the Model 2001-TCSCAN, a 9-channel, thermo-
couple card with a built-in cold junction. For external scanning, the Model 2000 is com-
patible with Keithley's Model 7001 and 7002 switch matrices and cards.
•
Programming languages and remote interfaces — The Model 2000 offers three program-
ming language choices (SCPI, Keithley Models 196/199, and Fluke 8840A/8842A) and
two remote interface ports (IEEE-488/GPIB and RS-232C).
•
Reading and setup storage — Up to 1024 readings and two setups (user and factory de-
faults) can be stored and recalled.
•
Closed-cover calibration — The instrument can be calibrated either from the front panel
or remote interface.
1-2
General Information
Содержание 2000
Страница 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 13: ...1 General Information...
Страница 19: ...2 Basic Measurements...
Страница 54: ...2 36 Basic Measurements...
Страница 55: ...3 Measurement Options...
Страница 87: ...4 Remote Operation...
Страница 139: ...5 SCPI CommandReference...
Страница 215: ...A Specifications...
Страница 226: ...A 12 Specifications...
Страница 227: ...B Statusand ErrorMessages...
Страница 233: ...C Example Programs...
Страница 246: ...C 14 Example Programs...
Страница 247: ...D Models196 199and 8840A 8842ACommands...
Страница 255: ...E IEEE 488 BusOverview...
Страница 271: ...F IEEE 488andSCPI ConformanceInformation...