![KCI V.A.C.Via v.a.c.therapy Negative Pressure Wound Therapy System Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/kci/v-a-c-via-v-a-c-therapy-negative-pressure-wound-therapy-system/v-a-c-via-v-a-c-therapy-negative-pressure-wound-therapy-system_instructions-for-use-manual_659637025.webp)
17
Guidance and manufacturer’s declaration - electromagnetic immunity
The
V.A.C.Via
™
Therapy Unit is intended for use in the electromagnetic environment specified below. The customer or user of the V
.A.C.Via
™
Therapy Unit should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV Contact
±
8 kV Air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV
for power supply lines
±
1 kV
for input/output lines
±
2kV
for power supply lines
±
1kV
for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±
1 kV line to line
±
2 kV line to earth
±
1 kV line to line
±
2 kV line to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<
5% X* for .5 cycle
<
40% X* for 5 cycle
<
70% X* for 25 cycle
<
5% X* for 5 seconds
<
5% X* for .5 cycle
<
40% X* for 5 cycle
<
70% X* for 25 cycle
<
5% X* for 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the
V.A.C.Via
™
Therapy Unit requires continued
operation during power mains interruptions, it
is recommended that the
V.A.C.Via
™
Therapy
Unit be powered from an uninterruptible power
supply or battery.
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: X* is the a.c. mains voltage prior to application of the test level.
Electromagnetic Compatibility:
Electromagnetic Interference - Although this equipment conforms with the intent of the directive
89/336/EEC in relation to Electromagnetic Compatibility (EMC), all electrical equipment may
produce interference. If interference is suspected, move equipment away from sensitive devices or
contact the manufacturer.
The Essential Performance requirements of the Model 60300 are to maintain 1/- 15
mmHg (30 second average) vacuum pressure with no false alarms.
Portable and mobile RF communications equipment can effect medical electrical equipment.
Radios, cell phones and similar devices may affect this equipment and should be kept at least 6.5
feet (2 meters) away from the equipment.
Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information in the following tables.
The following tables document compliance levels and guidance from the IEC 60601-1-2 2007
Standard, for the electromagnetic environment in which the V
.A.C.Via
™
Therapy Unit should be
used in a clinical environment. The V
.A.C.Via
™
Therapy Unit also meets the intent of the draft
guidance for Electromagnetic Compatibility related to use in the home care environment (IEC
60601-1-11 2007-11-30 draft).
The V
.A.C.Via
™
Therapy Unit was tested under model 60300 for EMC.
Guidance and manufacturer’s declaration - electromagnetic emissions
The V
.A.C.Via
™
Therapy Unit is intended for use in the electromagnetic environment specified below. The customer or user of the V
.A.C.Via
™
Therapy Unit should assure that it is used in such an environment.
Emission
test
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1
The V
.A.C.Via
™
Therapy Unit uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic
emissions
IEC 61000-3-2
Class B
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
complies
The VACVIA
™
is suitable for use in all establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Содержание V.A.C.Via v.a.c.therapy Negative Pressure Wound Therapy System
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 30: ......
Страница 32: ......
Страница 58: ......
Страница 60: ......
Страница 84: ......
Страница 86: ......
Страница 88: ......
Страница 114: ......
Страница 116: ......
Страница 140: ......
Страница 142: ......
Страница 144: ......
Страница 170: ......
Страница 172: ......
Страница 194: ......
Страница 196: ......
Страница 218: ......
Страница 220: ......
Страница 246: ......
Страница 248: ......
Страница 270: ......
Страница 272: ......
Страница 294: ......
Страница 296: ......
Страница 318: ......
Страница 320: ......
Страница 343: ......
Страница 344: ......