![KCI V.A.C.Via v.a.c.therapy Negative Pressure Wound Therapy System Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/kci/v-a-c-via-v-a-c-therapy-negative-pressure-wound-therapy-system/v-a-c-via-v-a-c-therapy-negative-pressure-wound-therapy-system_instructions-for-use-manual_659637022.webp)
14
The therapy unit will sound audible and display visual alarms as follows:
blOCkAGE AlARM
• A solid yellow LED above the blockage symbol will turn on.
• Audible blockage alarm will sound two beeps repeating every 15 seconds.
• When the blockage condition is resolved, audible and visual alarms will
turn off.
To Correct Alarm
Check for a full canister.
Check for kinked tubing.
Ensure tubing clamp is open.
lEAk AlARM
• A solid yellow LED above the leak symbol will turn on.
• Audible leak alarm will sound two beeps repeating every 15 seconds.
• When the leak condition is corrected, audible and visual alarms will turn off.
To Correct Alarm
See
Leaks
section in these instructions.
lOW bATTERY lEVEl AlARM
• A solid yellow LED on the battery level indicator will turn on.
• Alarm will sound two beeps repeating every four minutes
• A low battery alarm indicates approximately two hours of therapy remain;
charge batteries IMMEDIATELY to prevent disruption of therapy.
• When battery is charged, audible and visual alarms will turn off.
To Correct Alarm
Charge battery; see battery charge section in these instructions.
SYSTEM fAUlT AlARM
• All LEDs will turn on and flash.
• Two beeps sound, repeating every 15 seconds.
To Correct Alarm
Power the therapy unit off and then on again. If alarm continues contact the
treating physician or caregiver.
THERAPY ENDED
• A solid yellow LED at the top of the Therapy Life Indicator will turn on.
• The therapy unit will sound eight beeps, followed by a continuous beep for
five seconds, then the therapy unit will turn off.
• Notify the treating physician or caregiver.
If the therapy unit has completed the seven day therapy and has timed out, and an attempt is made to turn
the therapy unit on, the therapy unit will sound an audible alarm for three seconds then shut off.
WARNING:
Never leave a V.A.C.
®
Dressing in place without active V.A.C.
®
Therapy for more than two
hours. If therapy is off for more than two hours, remove the old dressing. Either apply a new V.A.C.
®
Dressing from an unopened sterile package and restart V.A.C.
®
Therapy; or apply an alternative dressing at
the direction of the treating physician or caregiver.
Содержание V.A.C.Via v.a.c.therapy Negative Pressure Wound Therapy System
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 30: ......
Страница 32: ......
Страница 58: ......
Страница 60: ......
Страница 84: ......
Страница 86: ......
Страница 88: ......
Страница 114: ......
Страница 116: ......
Страница 140: ......
Страница 142: ......
Страница 144: ......
Страница 170: ......
Страница 172: ......
Страница 194: ......
Страница 196: ......
Страница 218: ......
Страница 220: ......
Страница 246: ......
Страница 248: ......
Страница 270: ......
Страница 272: ......
Страница 294: ......
Страница 296: ......
Страница 318: ......
Страница 320: ......
Страница 343: ......
Страница 344: ......