![KCI V.A.C.Via v.a.c.therapy Negative Pressure Wound Therapy System Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/kci/v-a-c-via-v-a-c-therapy-negative-pressure-wound-therapy-system/v-a-c-via-v-a-c-therapy-negative-pressure-wound-therapy-system_instructions-for-use-manual_659637020.webp)
12
WARNING:
Never leave a V.A.C.
®
Dressing in place without active V.A.C.
®
Therapy for more than two
hours. If therapy is off for more than two hours, remove the old dressing and irrigate the wound. Either
apply a new V.A.C.
®
Dressing from an unopened sterile package and restart V.A.C.
®
Therapy; or apply an
alternative dressing at the direction of the treating physician or caregiver.
DRESSING REMOVAl
NOTE:
If dressing is lifted to observe wound, do not re-adhere the same dressing, a new dressing must
be applied.
1. Turn therapy unit off by pressing and holding the On / Off button for three seconds.
2. Gently stretch the drape horizontally to release the adhesive from the skin.
3. Do not peel vertically. Gently remove the drape from the wound.
4. Clean any residual adhesive with alcohol swab.
If a new dressing is to be applied:
1. Ensure that area is clean, using an alcohol swab or antiseptic wipe.
2. Allow skin to completely dry before applying.
3. Follow Dressing Application instructions, page 6.
CANISTER REMOVAl AND REPlACEMENT
1. Turn therapy off.
2. Slide dressing tubing clamp close to where tubing plugs into canister. Close clamp.
3. Unplug tubing from canister tubing ports.
4. Remove therapy unit from carrying case, if in use.
5. Depress tab on canister to remove used canister from therapy unit (Fig. 14).
6. Install new canister (see
Canister Installation
section in these instructions).
7. Return therapy unit to carrying case if desired.
8. Reattach dressing tubing to canister tubing ports.
9. Release tubing clamp.
10. Turn therapy on.
NOTE:
Dispose of used canister according to institution and local environmental regulations.
lEAkS
When the therapy unit detects a significant leak, a visual and audible leak alarm will activate (See also
the
Alarms
section of these instructions). When 125 mmHg is selected, the leak alarm will activate at
pressures of 75 mmHg or less. When 75 mmHg is selected, the leak alarm will activate at pressures of
50 mmHg or less.
1. During a leak condition, a solid yellow LED above the leak symbol will turn on. The system will sound
two beeps repeating every 15 seconds. The therapy unit will continue the alarm until condition is
corrected.
WARNING:
Never leave a V.A.C.
®
Dressing in place without active V.A.C.
®
Therapy for more than
two hours. If therapy is off for more than two hours, remove the old dressing. Either apply a new
V.A.C.
®
Dressing from an unopened sterile package and restart V.A.C.
®
Therapy; or apply an alternative
dressing at the direction of the treating physician or caregiver.
2. When leak condition has been corrected, audible alarms will stop, and visual alarms will turn off.
NOTE:
The green LED next to the therapy pressure selector button, 75 mmHg or 125 mmHg, will
remain on while the therapy system cycles through the steps above.
NOTE:
When 75 mmHg is selected, the operating range is 50 mmHg to 75 mmHg. When 125 mmHg
is selected, the operating range is 75 mmHg to 125 mmHg.
Содержание V.A.C.Via v.a.c.therapy Negative Pressure Wound Therapy System
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 30: ......
Страница 32: ......
Страница 58: ......
Страница 60: ......
Страница 84: ......
Страница 86: ......
Страница 88: ......
Страница 114: ......
Страница 116: ......
Страница 140: ......
Страница 142: ......
Страница 144: ......
Страница 170: ......
Страница 172: ......
Страница 194: ......
Страница 196: ......
Страница 218: ......
Страница 220: ......
Страница 246: ......
Страница 248: ......
Страница 270: ......
Страница 272: ......
Страница 294: ......
Страница 296: ......
Страница 318: ......
Страница 320: ......
Страница 343: ......
Страница 344: ......