1. With the ActiV.A.C.
®
Therapy Unit still powered on with the Clinician
Home Screen showing, press ‘?’, press ‘About’, press and hold the
‘Wrench Icon’
until the ‘Service Access’ screen appears.
2. Enter the code 772, press ‘OK’.
3. On the ‘Service’ screen, Press and hold ‘Patient Defaults (Press/Hold)’
until ‘Events Cleared (P)’ appears on the screen.
4. Press ‘Exit’.
5. Press ‘Cancel’.
6. Press ‘Exit’ to return to the Clinician Home Screen.
7. Power the unit off by holding down the Power On/Off button for a
minimum of two seconds.
8. Continue on to ‘Recharge Battery’.
Recharge Battery
The purpose of this procedure is to ensure the ActiV.A.C.
®
Therapy Unit is delivered to a patient placement with a fully
charged battery. Depending on the initial battery charge level, this process may take up to six hours.
1. With the ActiV.A.C.
®
Therapy Unit still powered off, set the unit in a secure place where it can be left for a minimum
of six hours.
2. Plug the power supply cord into the ActiV.A.C.
®
unit.
3. Plug the mains power cord into the external power supply ‘brick’.
4. Plug the mains power cord into a grounded AC mains wall socket.
5. Verify the Battery Charging Indicator LED is lit.
An amber light means the unit is charging the battery. A green light means the battery is fully charged.
6. Once the Battery Charging Indicator LED is green, continue on to ‘Preparation for Transport and Patient Use’.
Service
mm/dd/yy Time Event
asdf
12/06/09
15:48 Unit
On
12/06/09
15:55 Dressing
Changed,
4
12/06/09
15:57 Therapy
On
12/06/09
22:27 Canister
Full
12/06/09
22:27 Therapy
Off
12/06/09
22:31 Canister
Changed
12/06/09
22:32 Therapy
On
12/07/09
07:21 Auto
Therapy
Off
Leak
12/07/09
07:42 Therapy
On
12/07/09
14:26 Canister
Full
12/07/09
18:17 Therapy
Off
12/15/09
07:48 Events
Cleared
(P)
Exit
+
_
Next
Patient Defaults (Press/Hold)