![KCI ActiV.A.C. Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/kci/activ-a-c/activ-a-c_owners-service-manual_3774376012.webp)
10
Always use a USB Flash Drive. Do not plug a USB PC cable
into the ActiV.A.C.
®
Therapy Unit.
7. Insert a USB Flash Drive into the USB Data Port.
8. Press ‘Next’.
9. Wait while the unit downloads the data to the Flash Drive.
10. When the download is complete, Press ‘OK’.
• If System Errors 2 through 6 occur, power the unit off and back on and try again.
• System Errors 7 and 8 are cleared by resetting patient defaults, see
Final Settings
section.
• If the error recurs, contact KCI (see page 28).
• If System Error 1 occurs, contact KCI (see page 28).
11. Press ‘Exit’ to return to the Clinician Home Screen.
12. Remove the USB Flash Drive from the ActiV.A.C.
®
unit.
13. Connect the USB Flash Drive to an external computer.
The ActiV.A.C.
®
unit exports a file folder that uses the unit serial number as its name. The folder
contains a Comma Separated Values (CSV) file named by date, YYYYMMDD.csv.
14. For the initial save operation, save the entire file folder to a secure computer.
15. For subsequent data exports from the same ActiV.A.C.
®
unit, save the new dated CSV file to the same secure
computer into the existing folder named with the ActiV.A.C.
®
unit serial number.
16. With the Data Download complete, continue on to ‘Battery Check’.
Battery Check
The purpose of this procedure is to determine if the battery needs changing.
1. Locate the Battery History label located on the bottom of the unit.
2. Verify the battery change date on the battery history label.
3. If the battery is within the date range, skip ahead to ‘Pressure Checks’; otherwise proceed to the ‘Change Battery’
procedure.
Содержание ActiV.A.C.
Страница 2: ......
Страница 32: ...Manufactured for KCI USA Inc 12930 IH10 West San Antonio Texas 78249 USA www acelity com...