14
6. After acknowledging the warning screen, ensure the ActiV.A.C.
®
unit is
in Clinician Mode by observing the text in the upper right hand corner
of the screen, as illustrated.
7. If not in Clinician Mode, press ‘?’, press ‘Clinician Mode’, press and
hold ‘OK’ until the screen changes to Clinician Mode.
Clinician Mode
6 Aug 2009
15:06
25 mmHg Test
1. Press ‘Therapy’.
2. Press ‘Next’.
3. Press ‘Settings’.
4. Press ‘Pressure’ and use the arrow keys to set the pressure to 25.
5. Press ‘Exit two (2) times.
6. On the Confirm screen, ensure that settings are Pressure: 25 mmHg;
Mode: Continuous; Intensity: Low. If these settings are not correct,
press ‘Back’ to change the incorrect setting.
Continuous
Therapy Off
0 mmHg
Pressure
Exit
+
_
25 mmHG
7. Press ‘OK’.
8. Start therapy by pressing ‘On/Off’, the ‘Therapy Start’ screen will
appear.
Continuous
Therapy Off
0 mmHg
Confirm
OK
Back
Pressure:
25 mmHg
Mode:
Continuous
Intensity: Low
9. Adjust the needle valve so that the Leak Rate bar graph indicator on
the ‘Therapy Start’ screen indicates green and is below the dividing line
on the bar graph, see illustration.
10. Observe the digital manometer and the displayed pressure on the
ActiV.A.C.
®
Touch Screen.
11. The digital manometer should stabilize and indicate the pressure
displayed touch screen, ± 10 mmHg.
12. Press ‘Exit’.
13. Press ‘Therapy’.
14. Press ‘Next’.
15. Press ‘Settings’.
Continuous
Therapy On
25 mmHg
Log
Seal Audio
Therapy Start
Low
High
Leak Rate
Exit
125 mmHg Test
1. Press ‘Pressure’ and use the arrow keys to set the pressure to 125.
2. Press ‘Exit two (2) times.
Continuous
Therapy Off
0 mmHg
Pressure
Exit
+
_
125 mmHG