16
5. Press ‘Therapy’.
6. Press ‘Next’.
7. Press ‘Seal Check™’.
8. Adjust the needle valve so that the Leak Rate bar graph indicator on
the ‘Therapy Start’ screen indicates green and is below the dividing line
on the bar graph, see illustration.
9. Observe the digital manometer and the displayed pressure on the
ActiV.A.C.
®
Touch Screen.
10. The digital manometer should stabilize and indicate the pressure
displayed touch screen, ± 10 mmHg.
11. Press ‘Exit’.
12. Press ‘Settings’.
Continuous
Therapy On
200 mmHg
Seal Audio
Seal Check™
Low
High
Leak Rate
Exit
125 mmHg Test
1. Press ‘Pressure’ and use the arrow keys to set the pressure to 125.
2. Press ‘Exit two (2) times.
Continuous
Therapy Off
0 mmHg
Pressure
Exit
+
_
125 mmHG
3. On the Confirm screen, ensure that settings are Pressure: 125 mmHg;
Mode: Continuous; Intensity: Low.
4. Press ‘OK’.
5. Observe the digital manometer and the displayed pressure on the
ActiV.A.C.
®
Touch Screen.
6. The digital manometer should stabilize and indicate the pressure
displayed touch screen, ± 10 mmHg, within 60 seconds.
7. If the ActiV.A.C.
®
Therapy Unit passes all four (4) pressure change
procedures, remove the test fixture tubing and reinstall the Silicone
Canister Interface Seals on the unit.
8. Should the ActiV.A.C.
®
Therapy Unit fail any one of these pressure
changes by not displaying the correct pressure within 60 seconds,
contact KCI (see page 28).
9. With the Pressure Check complete, continue on to ‘Charge Battery and
Six Hour Verification Test’.
Continuous
Therapy Off
0 mmHg
Confirm
OK
Back
Pressure:
125 mmHg
Mode:
Continuous
Intensity: Low