![KCI ActiV.A.C. Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/kci/activ-a-c/activ-a-c_owners-service-manual_3774376014.webp)
12
6. Disconnect the battery from the unit.
7. Remove the old battery and dispose of it properly in accordance with
local, state and federal regulations for Lithium-ion (Li-ion) batteries.
The ActiV.A.C.
®
Therapy Unit battery must be replaced only
with batteries supplied by KCI. Use of any other battery may
compromise safety and/or cause erratic operation.
8. Obtain a new battery from KCI (KCI P/N 340079).
9. Obtain a new Battery History Label (KCI P/N 340168).
Shelf life of the battery is 12 months from battery
manufacture date. Verify that the battery manufacture
date is within 12 months of the date of installation. If the
manufacture date is outside 12 months, contact KCI (see page
28) for a replacement battery.
10. Calculate the next ‘Battery Change Due’ date as 18 months from the
installation date.
11. Record the required information on the Battery History label.
12. Slide new battery about halfway into unit until connectors meet. Plug
connectors together and ensure locking tabs are fully engaged.
13. Fit the battery into the housing ensuring the wiring harness is on the
front side of the battery and
NOT
behind and is otherwise not tangled
or obstructing fitting the battery fully into the battery compartment.
14. Press the Power On/Off switch to ensure the unit powers on properly.
15. Press the Power On/Off switch to turn the unit off.
16. Fit the battery cover, ensuring that it easily fits and does not need to
be forced into place. If force is required, the battery is not properly
installed. Double check that the wiring harness is not preventing the
battery from fitting fully into the battery compartment.
17. Install the two (2) screws tightly.
18. Install the new Battery History Label across the battery cover.
19. With the Battery Change complete, continue on to ‘Pressure Check’.
Screws
Battery History Label
Pressure Checks
The purpose of this procedure is to ensure the therapy unit pump responds to pressure setting adjustments.
The ActiV.A.C.
®
Therapy Unit MUST be connected to the ActiV.A.C.
®
power supply during this procedure.
The needle valve simulates the permeability of a V.A.C.
®
Drape.
The PVC test fixture and hygroscopic filter simulates the ActiV.A.C.
®
canister.
Set-up
1. Remove the Silicone Canister Interface Seals from the unit.
Seals Removed
Содержание ActiV.A.C.
Страница 2: ......