Inspect Unit for Damage
The purpose of this procedure is to inspect the ActiV.A.C.
®
unit for damage
using the following criteria:
1. Examine the unit top, bottom and sides for:
• Cracks of any size.
• Holes of any size that expose internal components.
2. Examine the canister attachment area for:
• Cracks of any size
• Damage to the canister seals
• Missing or damaged canister stabilization grommets
Service Procedures
The following service procedures are used to verify that the ActiV.A.C.
®
Therapy Unit is functioning properly and that the
battery is fully charged. Once these procedures are complete, the unit will be ready for patient use. These procedures
should be performed
as listed and in order:
1. When the unit is first received from KCI, as part of the initial inspection (starting at ‘Inspect Unit for Damage’).
2. Between each patient use.
All service should occur in a clean area that is protected from contamination from unclean units or
other potential contamination sources. Use PPE as appropriate or as specified in local protocols.
Parts and Equipment Needed
• ActiV.A.C.
®
Canister KCI P/N M8275058, two (2) each
• These are normal ActiV.A.C.
®
canisters that can be reused for up to 6 months as a test canister. KCI recommends
marking the canister with ‘Test Canister-Not For Human Use’ and the in-use date.
• Tubing Cap, KCI P/N M6275069, one (1) each. Note: A tubing cap is attached to the canister tubing as delivered.
• ActiV.A.C.
®
Therapy System Required Service Record form, KCI P/N 340141, one (1) each
• Canister Seal, KCI P/N 340040, two (2) each (if required)
• Exhaust Filter, KCI P/N 340037, one (1) each
• ActiV.A.C.
®
Fabric Carrying Case KCI P/N 340122, one (1) each
• ActiV.A.C.
®
Pressure Accuracy Accessory, KCI P/N 340145, one (1) each
• Calibrated Digital Manometer, customer supplied, one (1) each. Manometer accuracy to be within 2.5 mmHg.
Canister Seals
Canister Stabilization Grommets
Battery
History Label
Battery Charging/
Power Connection
USB Data Port
3. Examine the battery charging/power connection and USB data port to
ensure they are clear with no visible bent pins or other obstructions.
4. Examine the touch screen area and ensure it is secure in the case.
5. Check for excessive screen scuffing, scratches and/or cracks that could
obscure screen elements.
Deep or extensive screen scratches or cracks can be the source
of intermittent erratic therapy unit operation.