Charge Battery and Six Hour Unit Verification Test
The purpose of this procedure is to ensure the ActiV.A.C.
®
unit is fully functional and the battery is fully charged. This six
hour test is a minimum time. The process can be extended overnight or over a weekend if necessary.
1. Set the unit in a secure place where it can be left for a minimum of six hours. Charging can extend longer than six
hours if necessary.
2. Plug the power supply cord into the ActiV.A.C.
®
unit.
3. Plug the mains power cord into the external power supply ‘brick’.
4. Plug the mains power cord into a grounded AC mains wall socket.
5. Verify the Battery Charging Indicator LED is lit.
Battery charging in-process
Battery charging complete
An amber light means the battery is charging. A green light
means the battery is fully charged.
6. Ensure the unit is powered on. If necessary, hold down the Power On/
Off button for a minimum of two seconds to power the unit on.
7. Once the unit has completed its power on self test, verify the ‘plug’
icon is present on the lower left of the touch screen.
Continuous
Therapy Off
0mmHg
On/Off
Clinician Mode
6 Aug 2009
15:06
Utilities
Therapy
‘Plug’ Icon
Continuous
Therapy Off
0mmHg
On/Off
Clinician Mode
6 Aug 2009
15:06
Utilities
Therapy
‘Battery’ Icon
8. Unplug the power supply cord from the ActiV.A.C.
®
unit. Verify the
‘battery’ icon is present on the lower left of the touch screen.