Chapter 2. The Validator Hardware
2.7 Setting the Validator Time and Date
When the Console first establishes communication with the Validator, the current time and
date are automatically transferred to the attached Validator. No action on your part is
required. Refer to your Windows documentation for instructions for changing the date and
time.
2.8 Preventive Maintenance
2.8.1 Fuse Replacement
The Validator is equipped with a 250V T 4A fuse. The fuse is located in the line filter along
with the power cord. To replace the fuse:
1.
Turn the power off.
2.
Disconnect the power cord from the line filter.
3.
Using a small, slotted screwdriver (provided with the SIMs), pry open the fuse
holder from the bottom tab.
4.
Slide the fuse holder out and replace the fuse with a 250V T 4A replacement fuse.
2.8.2 Calibration
Each Validator is fully calibrated before shipment. The equipment normally requires no
further adjustment or calibration during installation. Under normal operating conditions, it is
recommended you calibrate the Validator once a year to maintain peak system accuracy. If
you use a Validator for high accuracy measurements, or if you replace or add circuit boards,
you may need to certify the calibration more often.
Kaye Validator User’s Manual
35
Содержание Validator AVS
Страница 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Страница 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Страница 3: ...No content intended for this page Kaye Validator User s Manual...
Страница 104: ...Chapter 9 The Qualification Study 9 2 1 Select a Setup File cont Kaye Validator User s Manual 97...
Страница 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Страница 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Страница 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Страница 167: ...No content intended for this page...