Chapter 2. The Validator Hardware
2.3.1b Connecting a Contact
You can connect dry contact inputs directly to the SIM, as shown in Figure 13 below. Dry
contact inputs can be used to make a time notation in the data file (to mark start exposure,
stop exposure, start of qualification, end of qualification) and to mark any events that occur
during the qualification study. You can also wire a dry contact to your vessel's PLC to detect
status output, and use it to automatically mark the start and end of the exposure cycle during
the qualification study.
+
– + – + – + – + – + –
1
2
3
4
5
6
Figure 13: Connecting Dry Contact Inputs
2.3.1c Connecting a Voltage Input
Connect a voltage input (up to 10 VDC) to the connectors, as shown in Figure 14 below.
Always connect the positive (+) lead to the positive (+) connector and the negative (-) lead
to the negative (-) connector.
+
– + – + – + – + – + –
1
2
3
4
5
6
Voltage
+
Transducer
–
Figure 14: Voltage Input Connection
Kaye Validator User’s Manual
29
Содержание Validator AVS
Страница 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Страница 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Страница 3: ...No content intended for this page Kaye Validator User s Manual...
Страница 104: ...Chapter 9 The Qualification Study 9 2 1 Select a Setup File cont Kaye Validator User s Manual 97...
Страница 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Страница 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Страница 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Страница 167: ...No content intended for this page...