Chapter 6. Defining Study Setups
6.8 Specifying Calibration Parameters (cont.)
Set the calibration parameters by specifying:
Temperature setpoints for calibration. You can specify a low setpoint, a high setpoint and a
check setpoint. If you specify both low setpoint and high setpoint, the high setpoint must be
at least 2° higher than the low setpoint. You must select either a low setpoint or a high
setpoint for calibration.
Temperature setpoints for calibration verification to verify that each sensor is still within your
calibration criteria. You must select at least one setpoint for calibration verification.
Sensor and IRTD stability in terms of temperature variation over a fixed time period. Stability
is the amount of change that is acceptable over the fixed time period.
Deviation criteria for uncalibrated temperature sensors, and deviation criteria for calibrated
temperature sensors over a fixed time period. Deviation is the difference in temperature
between the sensor values and the temperature standard.
Kaye Validator User’s Manual
71
Содержание Validator AVS
Страница 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Страница 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Страница 3: ...No content intended for this page Kaye Validator User s Manual...
Страница 104: ...Chapter 9 The Qualification Study 9 2 1 Select a Setup File cont Kaye Validator User s Manual 97...
Страница 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Страница 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Страница 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Страница 167: ...No content intended for this page...