Chapter 2. The Validator Hardware
2.5
Sensor and IRTD Installation (cont.)
CTR -80
To insert thermocouple sensors into the CTR -80:
•
Loosen the sliding clamp mechanism on a thermocouple well, insert the thermocouple
sensors 7.5 inches into the well, slide the clamp into place, and tighten the clamp to hold
the sensors in place. When properly installed, the sensors will protrude approximately
0.5 inches beyond the end of the thermocouple wells. For 22 gauge wire, 12 PTFE
thermocouples fit into each well.
Note:
The access cover, which is 7.5 inches wide, can be used as a measuring device in
positioning the sensors.
To install the IRTD into the CTR -80:
•
Insert the IRTD into one of the two 12-inch nylon spacers. An IRTD with the standard
18-inch stem fits properly. If you have an older model IRTD with a 15-inch stem, shorten
the length of the nylon spacers accordingly.
Kaye Validator User’s Manual
33