9 Technical specifications
KaVo Scan eXam One
59
Recommended separation distances between portable and mobile RF
communications equipment and the eXam6.
The eXam6 is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the
user of the eXam6 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the eXam6 as recom-
mended below, according to the maximum output power of the com-
munications equipment.
Rated
maximum out-
put
power of trans-
mitter W
Separation distance according to frequency of transmitter m
150 kHz to
80 MHz
80 MHz to
800 MHz
800 MHz to
2,5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the rec-
ommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maxi-
mum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher fre-
quency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Approved:
Reviewed: Jalkanen Joni 2017-02-13 16:00
Reviewed
See PDM system to determine the status of this document. Printed out: 2017-03-17 10:35:17
D510414, 5