9 Technical specifications
54
KaVo Scan eXam One
9.5 Electromagnetic Compatibility
(EMC) tables
NOTICE!
Medical electrical equipment needs
special precautions regarding EMC and needs to be
installed according to EMC information.
NOTICE!
Mobile RF communications equipment
can effect medical electrical equipment
Guidance and manufacturer’s declaration - electromagnetic emissions
The eXam6 is intended for use in the electromagnetic environment specified
below. The customer or the user of the eXam6 should assure that it is used in
such an environment.
Emissions
Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11E
Group 1
The eXam6 uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The eXam6 is suitable for use in all establish-
ments, including domestic establishments
and those directly connected to the public
low-voltage power supply network that sup-
plies buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not
applicable
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Complies