9 Technical specifications
KaVo Scan eXam One
57
Guidance and manufacturer’s declaration – electromagnetic immunity
The eXam6 is intended for use in the electromagnetic environment
specified below. The customer or the user of the eXam6 should assure
that it is used in such an environment.
Immu-
nity
Test
IEC
60601
Test Lev-
el
Compli-
ance
Level
Electromagnetic environment - guid-
ance
Conducte
d RF IEC
61000-4-
6
Radiated
RF IEC
61000-4-
3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equip-
ment should be used no closer to any part of the
eXam6,
including cables, than the recommended separa-
tion
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended Separation Distance:
where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the recommend-
ed separation distance in metres (m). Field
strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,
a
should
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
80 MHz to 800 MHz
800 MHz to 2.5 GHz