9 Technical specifications
KaVo Scan eXam One
55
Guidance and manufacturer’s declaration – electromagnetic immunity
The eXam6 is intended for use in the electromagnetic environment specified
below. The customer or the user of the eXam6 should assure that it is used in
such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environ-
ment - guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humid-
ity should be at least 30%.
Electrical fast
transients/
bursts
IEC 61000-
4-4
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Surge
IEC 61000-
4-5
± 1 kV differ-
ential mode
line to line
± 1 kV differ-
ential mode
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Voltage dips,
short inter-
ruptions and
voltage vari-
ations on
power supply
lines
IEC 61000-
4-11
< 5%
U
T
(> 95% dip in
U
T
) for 0.5 cy-
cle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
)
for 5 sec
< 5%
U
T
(> 95% dip in
U
T
) for 0.5 cy-
cle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
)
for 5 sec
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If user of the eXam6 requires
continued operation during
power mains interruptions, it
is recommended that the
eXam6 be powered from an
uninterruptible power supply
or a battery.