6 Installation of the imaging plate system
40
KaVo Scan eXam One
unit. Just enter the IP address into the IP field
and then click
OK
to connect the PC to the unit
(it is not necessary to hold down the Start key
and click
Send to Scanner
with the other PCs
once the unit has already got an IP address).
6.4.3 EXPRESS Share
1.
PC
: If the unit is to be used with several PCs se-
lect the Use Multiconnect check box and select a
unique Workstation identifier number (between
1 and 4), for the PC being configured, from the
drop down list. Addition workstation informa-
tion, for example, user name, location etc, can
be entered into the field next to the work
station identifier number.
NOTICE!
If only one PC is connected to the unit
do not select the Use Multiconnect check box.
The Scanner Autorelease timeout is the length
of time that the unit remains reserved and un-
used by a PC before the PC automatically re-
leased the unit so that it can be used by another
PC in the system (the scanner can be reserved
in advance from another PC). The default setting
is 40 seconds. This can be changed by keying in
a new value.
2. Click OK to connect the PC to the unit.
3.
NOTICE!
An automatic technique automatically
locates the unit within the local area network
and connect the PC.
4. Repeat the above process for all the other PCs in
the network. Make sure that you give each PC a
different Workstation identifier.
5. Check the installation by starting image capture
using the imaging application software. If the
Use Multiconnect was selected the Workstation
identifier of the PC (1 - 4) being used appears on
the unit user interface.