9 Technical specifications
KaVo Scan eXam One
51
9.2 System requirements and
connections
Minimum requirements for the PC/laptop, network adapter and network
switch
PC / laptop network switch
Class I or Class II according IEC 60950
Network connection settings
10/100Mbs LAN
UDP/IP protocol traffic allowed
Traffic to UDP port 10000 allowed (unit UDP
port)
UDP broadcast traffic allowed
CAT6 Ethernet cable
DHCP server is recommended but not neces-
sary
Use
Use antivirus software.
Use firewall.
When LAN configuration is changed or devic-
es are added/removed, it may affect existing
devices in the LAN. Therefore keep in mind
that correct operation of the imaging system
needs to be checked after changes are
made.
When adding new devices to LAN, make sure
they all have unique IP address, otherwise
they may cause communication problems
with existing LAN devices. Place unit and PC
with imaging application software to same
subnet in LAN.
NOTICE!
Image is not transferred from unit to PC imaging application software in
case of connection lost during image processing. Image is stored in unit memory
until it has been transferred to PC. Unit cannot be turned off in that case. When
network is operational again, image is automatically transferred to imaging
application software. Do not disconnect unit PSU adapter before network is
operational and image has been transferred to imaging application software.
For more details of the hardware requirements running the imaging application
software please refer to the user manual of it.
Approved:
Reviewed: Jalkanen Joni 2017-02-13 16:00
Reviewed
See PDM system to determine the status of this document. Printed out: 2017-03-17 10:35:17
D510414, 5