3 Replacing and checking cleaning and germ reduction agents | 3.3 Check the OXYGENAL 6 concentration
Other concentrations can be tested using the following table:
H2O2 concentration
Measuring cup
Ratio water : treated
water
Value shown on test
strip
0.02 % (200 ppm)
standard concentration
40 ml
39:1
5 mg/l
0.05 % (500 ppm)
100 ml
99:1
5 mg/l
0.1 % (1,000 ppm)
40 ml
39:1
25 mg/l
0.2 % (2,000 ppm)
80 ml
79:1
25 mg/l
0.25 % (2,500 ppm) in-
tensive germ reduction
concentration
100 ml
99:1
25 mg/l
0.3 % (3,000 ppm)
120 ml
119:1
25 mg/l
45 / 48