![KaVo ESTETICA E70 Vision Скачать руководство пользователя страница 39](http://html2.mh-extra.com/html/kavo/estetica-e70-vision/estetica-e70-vision_care-instructions_3711872039.webp)
Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.9 Preparation of the amalgam separator
2.9 Preparation of the amalgam separator
The following systems can be used as options:
▪ "Dürr" Amalgam separator
▪ Preparation for external wet suction for connection to a central amalgam
separator
▪ "Dürr" separation kit
The correct and regular use of the cleaning programmes results in optimal ser-
vicing of the amalgam separation system.
Note
follow the instructions for use for the amalgam separator.
Replacing the amalgam container
▶
Turn off the treatment centre.
▶
Remove the cover.
▶
Replace the amalgam container according to the instructions for use of the
amalgam separator.
2.10 Reprocessing KaVo Screen One / KaVo Screen HD
Cleaning and disinfection
NOTICE
Non-colour-fast textiles.
Non-colour-fast textiles can cause permanent discoloration.
▶
Remove discolorations with water immediately.
NOTICE
Damage caused by spray disinfection.
Disinfectant gets into cracks and loosens the paint.
▶
Do not use spray disinfection.
▶
Only disinfect by wiping.
▶
Clean the surfaces with a soft cloth and water, and wipe-disinfect them with
an approved surface disinfectant.
Approved disinfectants:
▪ CaviCide / CaviWipes (Metrex)
39 / 48