Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
18 / 48
▶
If necessary, replace the turbine return air filter (
Mat. no. 0.200.3098
)
and clean the filter holder.
Note
The turbine return air filter is in the same position in the dentist element vari-
ant with S table.
2.4.7 Reprocessing of the kit for physiological saline
solution and surgical motor
The kit for the physiological saline solution and surgery motor are optional.
Note
The reprocessing steps for surgical motors with motor cable and for the
straight and contra-angle handpieces are described in the corresponding In-
structions for use.
Note
The reprocessing instructions have been validated by the manufacturer. Any
departure from the instructions provided must be checked by the repro-
cessing entity for efficacy and possible detrimental consequences.
Cleaning the kit for physiological saline solution
Manual cleaning
▶
Use a soft disposable cloth to wipe down all visible surfaces of the unit,
bottle holder, foot control surfaces, and connecting cables.