Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.7 Automatic cleaning programmes
ð
A progress bar indicates the cleaning steps.
S-table
▶
Lock the swing-arm into the front catching position to ensure that the hand-
piece is removed.
▶
In the case of swing-arms without a catching function, insert the swing-arm
lock (
Mat. no. 1.008.1511
) to ensure that the handpiece is removed.
▶
Grasp the three-function or multifunctional handpiece on the adapter
between hose and handpiece and take off the cannula and the gripping
sleeve.
▶
Place the three-function or multifunctional handpieces without cannulas or
gripping sleeves in the middle holders of the hygiene centre. Make sure to
place triple function handpieces in the grey holder and multifunctional hand-
pieces in the black holder of the hygiene centre.
▶
Place the handpiece hoses of the dentist element in the designated places
of the hygiene centre.
Note
All water regulation mechanisms for the motor, instruments, and multiflex
couplings must be open.
31 / 48