Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.7 Automatic cleaning programmes
36 / 48
ð
A progress bar indicates the cleaning steps:
▶
Wait for the cleaning programme to be completed.
ð
The following message is displayed after completion of the cleaning pro-
gramme:
▶
After completion of the "Weekly" cleaning programme, place hoses in hold-
ers.
Intensive germ reduction during extended periods of
non-use (> 4 weeks)
During extended periods of non-use, e.g. semester breaks at university,
"Weekly" cleaning must be run during the time of non-use. The germ reduction
liquid cleaner remains inside the system during the period of non-use.
Before the period of non-use
▶
Start the "Weekly" cleaning programme and switch it off during the active
phase after an acoustical signal within the dwell time of 30 minutes.
▶
To prevent damage due to hoses getting kinked, remove hoses from the
hygiene centre and place them in the dentist holder.
▶
Cover the treatment unit during extended periods of non-use.
After the period of non-use
▶
Turn on the central compressor.
▶
Switch on the water supply for the treatment unit after the system pressure
is achieved.
▶
Replace the handpiece hoses of the dentist and assistant element in the
places provided in the hygiene centre.
▶
Turn the treatment unit on.
ð
The unit automatically starts the remaining steps of the unfinished cycle.
▶
Allow the cycle to complete and then run a second full "weekly" cleaning.
ð
The unit is now ready for use.