Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
16 / 48
Drying the triple-function and multifunction handpiece
Manual drying
▶
Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic drying
Not applicable.
Care of the triple-function and multifunctional
handpiece
Requirement
The multifunctional handpiece has been reprocessed.
▶
Apply KaVo silicone grease (
Mat. no. 1.000.6403
) or KaVo Rota Spray 2
(
Mat. no. 0.411.7510
) to the O-rings of the joint between the gripping
sleeve and cannula. Use cotton buds for this purpose.
Sterilising the triple-function and multifunction
handpiece
Note
The grip sleeve and cannula can be sterilised.
▶
Hold the 3-way or multifunctional handpiece at the gripping sleeve and take
off the cannula with a slight twisting motion.
▶
Pull the gripping sleeve off the valve body.