Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.5 Preparation of the assistant unit
NOTICE
Damage to the camera due to improper disinfection.
Irreversible destruction of the camera.
▶
Disinfect with wipes only, do not use disinfectant sprays!
▶
Do not immerse the camera in liquids!
▶
Do not sterilise the camera in an autoclave!
▶
Do not store the camera in Steri pack!
▶
Do not sterilise the camera with hot air or in a washer disinfector!
▶
Switch main device switch off.
▶
Allow the camera to cool down.
▶
Remove the camera.
▶
Disinfect the surface using a soft wipe and disinfectant.
Approved disinfectants:
▪ Metrex CaviCide - manufacturer: Metrex
Note
The exterior of the camera and camera hose can be disinfected with chemical
disinfectants.
Comply with the manufacturer's instructions for use with regard to the applic-
ation range. and exposure time.
NOTICE
Material damage due to use of non-approved disinfectant.
Damage to the camera.
▶
Use approved disinfectants only.
2.5 Preparation of the assistant unit
2.5.1 Preparing the instrument holder
The instrument holder on the assistant unit is prepared in the same way as the
instrument holder on the dentist unit.
23 / 48