Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
2.4 Preparation of the dentist unit
2.4.1 Preparation of the handle and holder pad/ anti-
slip mat
Cleaning and disinfection of the handle and holder pad/
antislip mat (T table / S table)
▶
Press the push button and pull off the handle.
▶
Clean the handle under flowing water.
▶
Disinfect the white and grey handle by wiping or thermodisinfection (minor
discolouration may occur).
▶
Sterilise the holder pad and anti-slip mat by thermodisinfection.
Sterilisation of the handle and holder pad/ anti-slip mat
(T table / S table)
CAUTION
Damage caused by sterilisation.
▶
Do not sterilise the white standard handle.
▶
Sterilise the holder pad/ anti-slip mat.
The grey handle (
Mat. no. 1.002.4489
) can be sterilised (see symbol on the
handle).