6
Reconditioning methods according to EN ISO 17664
The listed instructions for cleaning and sterilising were validated as being suitable
by the medical device manufacturer for preparing a medical device. This usually
requires validation and routine monitoring of the procedure. Moreover, any devia‐
tion by the rehabilitating entity from the instructions provided should be carefully
analysed for impact on efficacy and possible consequences.
The following components must be reconditioned:
▪ Surface of the device
▪ Tips
CAUTION
Damage due to penetrating liquids.
Malfunctions from penetrated liquids.
▶ Do not allow any liquids to enter the device!
6.1
Preparations for cleaning
Note
The software must always be closed, before unplugging the DIAGNOcam.
▶ Remove tip.
6.2
Cleaning
Note
Do not use solvents or aggressive chemicals!
▶ Remove major soiling directly after soiling with a single-use paper towel.
6.3
Manual cleaning
▶ Clean all outer surfaces of the DIAGNOcam with a soft cloth and one of the
specified disinfectants.
▶ Clean the tips under running water (tap water quality, temperature: 30°C ±
5°C, flow rate: 2 l/min) for 30 seconds using a medium-hard toothbrush.
▶ If the window is soiled, clean it with isopropanol 70% and a Q-tip.
Instructions for use DIAGNOcam 2170
6 Reconditioning methods according to EN ISO 17664 | 6.1 Preparations for cleaning
54/64