International (EU)
For country-specific information on disposal, contact your dental supplier.
2.5
Safety instructions
WARNING
Danger of injury from electric current.
Electric shock.
▶ Stop working if the device becomes damaged.
▶ Connect the device only to a PC/laptop that is approved according to IEC
60950.
▶ Do not use the device on patients or place it near patients with the probe re‐
moved.
▶ Do not use the device after it has been dropped.
WARNING
Danger of suffocation.
Vomiting can be triggered by inserting the device too far.
Aspiration of vomit.
▶ Keep the device away from the patient's throat!
WARNING
Blinding hazard from invisible laser.
Eye damage.
▶ Do not point the device toward the eyes when the laser is active!
▶ Do not operate the device when the housing is damaged or opened.
▶ Do not use the handpiece on the patient when the tips are not attached.
▶ Do not look into the handpiece's aperture for the light when the tip is re‐
moved.
WARNING
Hazard from electromagnetic radiation.
Interference with other devices.
▶ Do not use the device on patients with pacemakers!
▶ Turn off devices situated in the treatment room that are sources of hazard‐
ous energy (such as x-ray machines, lasers, and rotating instruments)!
CAUTION
The product may be damaged by kinking or pinching the USB cable.
Irreversible breakage of the electrical lines in the USB cable.
▶ Do not pull on the USB cable!
Instructions for use DIAGNOcam 2170
2 Safety | 2.5 Safety instructions
8/64